View Item 
      •   R2KN Home
      • Politeknik Kesehatan Kemenkes
      • Politeknik Kesehatan Kementerian Kesehatan Semarang
      • Repository Politeknik Kesehatan Kemenkes Semarang
      • View Item
      •   R2KN Home
      • Politeknik Kesehatan Kemenkes
      • Politeknik Kesehatan Kementerian Kesehatan Semarang
      • Repository Politeknik Kesehatan Kemenkes Semarang
      • View Item
      JavaScript is disabled for your browser. Some features of this site may not work without it.

      UJI KINERJA PESAWAT MAGNETIC RESONANCE IMAGING MERK TOSHIBA VANTAGE ELAN 1,5 TESLA DI INSTALASI RADIOLOGI RUMAH SAKIT TK. III SLAMET RIYADI SURAKARTA

      Date
      2019
      Author
      Irchas Tika Arafati
      Metadata
      Show full item record
      Abstract
      MRI equipment at Radiology Installation of TK. III Slamet Riyadi Surakarta Hospital installed since September 2016 and has never done performance test. According to ACR (2015) MRI performance testing performed weekly and annually consisting of 7 parameters. This study aims to determine the results of performance testing of the MRI equipment brand Toshiba Vantage Elan 1.5 Tesla in Radiology Installation of TK. III Slamet Riyadi Surakarta Hospital and follow-up of the performance test.This type of research is quantitative with survey approach. Data were taken in April 2019 at Radiology Installation of TK. III Slamet Riyadi Surakarta Hospital. Research used an ACR phantom and head coil. This research tested seven parameters including geometric accuracy, high contrast spatial resolution, low contrast detectability, slice position accuracy, slice thickness accuracy, image intensity uniformity and percent signal ghosting. The data obtained is processed and analyzed in accordance with ACR international standard.Test result in 7 parameters are no difference in geometric accuracy about all x, y, both diagonals and z axis, high contrast spatial resolution can show resolution insert until second group, low contrast detectability is amount can show more than 9 spokes, slice position accuracy is amount 4 mm and 5 mm, slice thickness accuracy is amount 4,42 mm, image intensity uniformity is amount 88,47%, and percent signal ghosting is amount 0,85%. The results of performance test showed that the modality of MRI had still good. The results of testing 7 parameters are in accordance with ACR international standards. The resulting data can be used as baseline data for subsequent testing.Keywords: MRI of 1.5 Tesla, performance test, ACR phantom
      URI
      http://repository.poltekkes-smg.ac.id//index.php?p=show_detail&id=19932
      Collections
      • Repository Politeknik Kesehatan Kemenkes Semarang [19935]

      Copyright © Badan Litbangkes - Kementerian Kesehatan RI
      Jl. Percetakan Negara no.29, Jakarta Pusat 10560, Indonesia
      All rights reserved. 2019
      Contact Us | Send Feedback
        

       

      Browse

      All of R2KNCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects

      My Account

      LoginRegister

      Copyright © Badan Litbangkes - Kementerian Kesehatan RI
      Jl. Percetakan Negara no.29, Jakarta Pusat 10560, Indonesia
      All rights reserved. 2019
      Contact Us | Send Feedback